J 2019

The Combined Relative Uncertainty of Measurement Results by Prototype Semi-Automated Calorimetric Chamber

KUŠNEROVÁ, Milena; Jan VALÍČEK; Marta HARNIČÁROVÁ; Ján KMEC; Michal ŘEPKA et. al.

Basic information

Original name

The Combined Relative Uncertainty of Measurement Results by Prototype Semi-Automated Calorimetric Chamber

Name in Czech

Kombinovaná relativní nejistota výsledků měření prototypovou poloautomatickou kalorimetrickou komorou

Authors

KUŠNEROVÁ, Milena (203 Czech Republic, guarantor, belonging to the institution); Jan VALÍČEK (203 Czech Republic, belonging to the institution); Marta HARNIČÁROVÁ (203 Czech Republic, belonging to the institution); Ján KMEC (703 Slovakia, belonging to the institution); Michal ŘEPKA (203 Czech Republic, belonging to the institution); Roman DANEL (203 Czech Republic, belonging to the institution); Anton PANDA (703 Slovakia) and Zuzana PALKOVÁ (703 Slovakia, belonging to the institution)

Edition

MEASUREMENT SCIENCE REVIEW, POLAND, DE GRUYTER POLAND, 2019, 1335-8871

Other information

Language

English

Type of outcome

Article in a journal

Field of Study

20303 Thermodynamics

Country of publisher

Poland

Confidentiality degree

is not subject to a state or trade secret

References:

Impact factor

Impact factor: 0.900

RIV identification code

RIV/75081431:_____/19:00001548

Organization unit

Institute of Technology and Business in České Budějovice

UT WoS

000466984000003

Keywords (in Czech)

Kombinovaná relativní nejistota; krokové měření teploty; měrná tepelná kapacita; poloautomatická kalorimetrická komora

Keywords in English

"Combined relative uncertainty; step temperature measurement; specific thermal capacity; semi-automated calorimetric chamber "

Tags

Changed: 22/4/2020 14:11, Mgr. Milada Šanderová

Abstract

V originále

"The paper presents an evaluation of the combined relative uncertainty of the result of direct step temperature measurements aimed at evaluation of the indirect measurements of the specific thermal capacity of the heat insulating concrete by means of a pair of resistive cable thermometers fitted with Pt100 temperature sensors and integrated into a computer-controlled calorimetric chamber. In particular, it is a proposal of evaluation of the overall relative uncertainty of the measurement of partial temperatures measured in equidistant time steps, in a relatively wider time interval. In practice, the uncertainty of the result of step temperature measurements is most often declared only by the instrument uncertainty specified by the manufacturer. The exact evaluation of the result of the measurements of thermal and temperature material parameters measured by the calorimetric comparison method is required by the fact that the investigated samples are made of newly designed non-tabulated building materials and that the measurements are made by a prototype device."

In Czech

"Příspěvek prezentuje vyhodnocení kombinované relativní nejistoty výsledku přímého měření teplotních stupňů vyhodnocení nepřímých měření měrné tepelné kapacity tepelně izolačního betonu pomocí dvojice odporových kabelů teploměry vybavené teplotními čidly Pt100 a integrované do počítačově řízené kalorimetrické komory. Zejména to je návrh hodnocení celkové relativní nejistoty měření dílčích teplot měřených v ekvidistantních časových krocích, v relativně delší časový interval. V praxi je nejistota výsledku měření krokových teplot nejčastěji deklarována pouze pomocí nejistotu stanovenou výrobcem. Přesné vyhodnocení výsledku měření teploty a teploty Materiálové parametry měřené kalorimetrickou srovnávací metodou jsou vyžadovány tím, že zkoumané vzorky jsou vyrobeny nově konstruované nestrukturované stavební materiály a měření jsou prováděna prototypovým zařízením."

Files attached